• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. CARELINK; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. CARELINK; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2090
Device Problem Corroded (1131)
Patient Problem No Patient Involvement (2645)
Event Date 04/24/2019
Event Type  malfunction  
Manufacturer Narrative
Product event summary: manufacturer¿s analysis was unable to confirm the customer comments regarding an inability to load software on the programmer.It was found, however, that the programmer failed incoming functional tests due to excessive internal corrosion.The programmer was scrapped.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the programmer, originally returned due to the inability to load software after multiple attempts, subsequently tested out of specification during manufacturer¿s analysis.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
lisa robertson
8200 coral sea st ne
mounds view, MN 55112
7635262723
MDR Report Key8612869
MDR Text Key145147406
Report Number2182208-2019-00896
Device Sequence Number1
Product Code KRG
UDI-Device Identifier00613994581594
UDI-Public00613994581594
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 05/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/15/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2019
Date Manufacturer Received04/24/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/29/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-