The customer reported that the central nurse's station (cns) is not alarming v-tach and the org multiple patient receiver provides the signal analysis from the transmitter.
This is a use error issue as the appropriate leads were not used for arrythmia analysis and the recommendation was to use multi-read analysis or change trace 1 to ii lead and do single-read analysis.
This is not a device failure.
Nkc investigation: in single-lead analysis, some qrs complexes were not detected due to low amplitude, and the remaining qrs-detected beats were judged as the v-tach alarm did not occur.
The average heart rate decreased slightly to about 72 bpm, but it did not alarm because bradycardia was set to 50 bpm or less.
Therefore, it is determined that the v-tach alarm did not occur because trace 1 has low amplitude.
Recommendation: use multi-read analysis or change trace 1 to ii lead and do single-read analysis.
|