CONCORD MANUFACTURING 2008K HEMODIALYSIS SYS. OLC/DIASAFE PLUS DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 190371
Device Problems Thermal Decomposition of Device (1071); Melted (1385); Smoking (1585)
Patient Problem No Patient Involvement (2645)
Event Date 05/07/2019
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process. A supplemental mdr will be submitted upon completion of this activity.  .
 
Event Description
A biomedical technician (biomed) at a user facility reported that a fresenius 2008k hemodialysis (hd) machine had a burned and melted valve 33. The issue was found by the biomed during machine repair for leaks in heat disinfect mode. The biomed confirmed the valve also gave off smoke and a burning smell. A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction. The machine has approximately 22,000 hours of use and the valve was the original fresenius part on the machine. The biomed reported that there was no damage observed on any other components, or any other additional issues, associated with the burned and melted distribution board. The biomed replaced valve 33, which resolved the issue. The unit was returned to service at the user facility without issue and without reoccurrence of the event. The biomed confirmed no parts were available for return.
 
Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation. Additionally, no on-site evaluation was performed by a fresenius regional equipment specialist (res). A records review was performed on the reported serial number. An investigation of the device manufacturing records was conducted by the manufacturer. There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event. In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements. The investigation into the cause of the reported problem was not able to be confirmed. A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name2008K HEMODIALYSIS SYS. OLC/DIASAFE PLUS
Type of DeviceDIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
matthew amaral
920 winter st.
waltham, MA 02451
7816999758
MDR Report Key8621266
MDR Text Key145418841
Report Number2937457-2019-01548
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K994267
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 06/05/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received05/17/2019
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number190371
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Device Age MO
Event Location No Information
Date Manufacturer Received06/05/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/23/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-