Model Number N/A |
Device Problem
Device Contaminated During Manufacture or Shipping (2969)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/26/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported during surgery that the inner packaging had mold on it.Subsequently, the surgery was completed with a different device.No adverse events have been reported as a result of the malfunction.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay updated and additional information.Complaint sample was evaluated and the reported event was confirmed.Visual evaluation of the returned product identified that there is porous debris inside the sterile package and sterile pouch has scratches from product abrasion.Dhr was reviewed and no discrepancies were found.Investigation results concluded that the reported event was due to transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|