Catalog Number SGC0302
Device Problems Difficult to Insert (1316); Improper or Incorrect Procedure or Method (2017); Deformation Due to Compressive Stress (2889); Difficult to Advance (2920); Material Split, Cut or Torn (4008)
Patient Problem Perforation (2001)
Event Date 04/29/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4). During processing of this complaint, attempts were made to obtain complete event, patient and device information. (b)(4). The customer reported the device was discarded. Investigation is not yet complete. A follow up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report the torn soft tip, perforation and medical intervention. It was reported that this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4. The steerable guide catheter (sgc) was advanced into the femoral vein, but resistance was felt. Force was applied, but the sgc tip became bent and torn. The sgc was removed to be replaced; however, bleeding was observed at the puncture site. An unspecified covered stent was deployed to treat the perforation. The procedure was continued. One clip was implanted, reducing mr to 1-2. There was no clinically significant delay in the procedure. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMITRACLIP SYSTEM STEERABLE GUIDE CATHETER
Type of DeviceSTEERABLE GUIDE CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
MENLO PARK, CA REG# 3005070406
abbott vascular
26531 ynez rd.
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key8627761
MDR Text Key145638478
Report Number2024168-2019-03962
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Remedial Action Other
Type of Report Initial,Followup
Report Date 09/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/21/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date02/18/2020
Device Catalogue NumberSGC0302
Device Lot Number90216U157
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/23/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 05/21/2019 Patient Sequence Number: 1
-
-