Allegedly the patient was revised due to a fractured liner, femoral head pseudo-tumor and metallosis (left).Subject presented at clinical follow-up on (b)(6) 2015 with implant cracking/clicking, as well a some calcification or ceramic debris on the x-ray.Subsequent clinical follow-ups indicated increasing squeaking/noises form hip.Subject presented to emergency (b)(6) 2017 with hip issues, and was scheduled for revision on (b)(6) 2017.Intraoperatively it was found the liner and femoral head were fractured and a small pseudo-tumor resulting from metallosis was present.Acetabular and femoral components were removed and revised.(b)(4).
|