MAUDE Adverse Event Report: CONCORD MANUFACTURING LIBERTY SELECT CYCLER ASSY(NON-VALUATED); SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING LIBERTY SELECT CYCLER ASSY(NON-VALUATED); SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number RTLR180343
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/07/2019
Event Type  malfunction  
Manufacturer Narrative
Plant investigation: the actual device was returned to the manufacturer for physical evaluation.An exterior visual inspection of the returned cycler showed no signs of physical damage.Although there were visual indications of dried fluid within the cassette compartment, there were no burrs or sharp edges in cassette area that may have punctured a cassette membrane.Upon power up, the cycler touch screen test failed.When powering on the cycler, the ok, stop and up/down arrow push buttons illuminated, however the front panel touch screen remained dim.It was identified that the cause for the dim screen was due to an internal short on the transformer (t1) on the inverter board.The inverter board is located on the rear of the touch screen.A known good inverter board was installed, and the display became fully operational.The functioning inverter board was removed at the completion of the investigation.The mushroom head check passed.The cycler tested positive for glucose.A review of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformance during the manufacturing process.In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements.Upon completion of the evaluation, the reported issue was confirmed and the cause was determined to be an internal short of the transformer on the inverter board.The cycler was refurbished following the evaluation.
 
Event Description
A contact of a peritoneal dialysis (pd) patient called fresenius technical support to report the liberty cycler has a display brightness problem during step 10 of the treatment end screens.Patient contact stated that the power cord is loose from the back of the cycler and has to tape it to keep it in place.Patient contact stated that the screen is also dim and hard to see.Brightness was set to 9, was changed to 10, but after adjusting the power cord screen became dim again.Cycler is plugged directly into a three prong wall outlet, and shared with a mini-refrigerator.The fresenius technical support representative advised patient's contact that it is best to have cycler on its own outlet.The fresenius technical support representative issued a replacement cycler and advised the patient to discontinue using the machine.There was no patient harm or adverse event, and no medical intervention was required.Alternate treatment option is not available.Upon follow up, it was confirmed that no patient adverse effects were experienced and no medical intervention was required.The patient did not complete the treatment that day but resumed treatments on the new cycler with no further issue.The cycler was returned and upon physical evaluation of the cycler by the manufacturer, it was identified that the transformer on the inverter board had an internal short.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBERTY SELECT CYCLER ASSY(NON-VALUATED)
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
matthew amaral
920 winter st.
waltham, MA 02451
7816999758
MDR Report Key8645546
MDR Text Key146221005
Report Number2937457-2019-01678
Device Sequence Number1
Product Code FKX
UDI-Device Identifier00840861102068
UDI-Public00840861102068
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181108
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 05/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/28/2019
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberRTLR180343
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2019
Device Age MO
Date Manufacturer Received05/21/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/04/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
DELFLEX PD FLUID; LIBERTY CYCLER SET 
-
-