CONCORD MANUFACTURING LIBERTY SELECT CYCLER ASSY(NON-VALUATED); SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number RTLR180343
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Patient Involvement (2645)
Event Date 04/08/2019
Event Type  malfunction  
Manufacturer Narrative
Plant investigation: the actual device was returned to the manufacturer for physical evaluation.A visual inspection of the returned cycler exterior showed no sign of physical damage.The touch screen test failed - when powering on the cycler the ¿ok, stop, and up/down arrows push buttons¿ illuminated, however the front panel touch screen remained blank.An internal inspection of the cycler found transformer (t1) on the ¿inverter board¿ to have internal short.The inverter board is located on the rear of the touch screen.A known good inverter board was installed and the touch screen became operational.The functioning inverter board was removed from the touch screen at the completion of the investigation.There were no other discrepancies encountered in the internal inspection of the cycler.A review of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformance during the manufacturing process.In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements.Upon completion of the evaluation, the reported issue was confirmed and the cause was determined to be a shorted transformer on the inverter board.The cycler was refurbished following the evaluation.
 
Event Description
It was reported that the screen of a patient¿s liberty select cycler went blank prior to their dialysis (pd) treatment.At that point in time, the technical support representative advised the patient to discontinue use of the cycler and to notify their peritoneal dialysis registered nurse (pdrn) of the event.A replacement cycler was issued to the patient.It was reported that an alternate treatment option was available.The cycler was returned to the manufacturer and a replacement cycler was provided.Upon physical evaluation of the cycler by the manufacturer, it was identified that the transformer on the inverter board was shorted.
 
Event Description
It was reported that the screen of a patient¿s liberty select cycler went blank prior to their dialysis (pd) treatment.At that point in time, the technical support representative advised the peritoneal dialysis nurse (pdrn) that was assisting the patient to discontinue use of the cycler.A replacement cycler was issued to the patient.It was reported that an alternate treatment option was available.The cycler was returned to the manufacturer and a replacement cycler was provided.Upon physical evaluation of the cycler by the manufacturer, it was identified that the transformer on the inverter board was shorted.
 
Manufacturer Narrative
Corrected plant investigation: the actual device was returned to the manufacturer for physical evaluation.A visual inspection of the returned cycler exterior showed no sign of physical damage.The touch screen test failed - when powering on the cycler the ¿ok, stop, and up/down arrows push buttons¿ illuminated, however the front panel touch screen remained blank.An internal inspection of the cycler found transformer (t1) on the ¿inverter board¿ to have internal short.The inverter board is located on the rear of the touch screen.A known good inverter board was installed and the touch screen became operational.The functioning inverter board was removed from the touch screen at the completion of the investigation.There were no other discrepancies encountered in the internal inspection of the cycler.A review of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements.Upon completion of the evaluation, the reported issue was confirmed and the cause was determined to be a shorted transformer on the inverter board.The cycler was refurbished following the evaluation.
 
Manufacturer Narrative
Removed consumer from report source plant investigation: the actual device was returned to the manufacturer for physical evaluation.A visual inspection of the returned cycler exterior showed no sign of physical damage.The touch screen test failed - when powering on the cycler the ¿ok, stop, and up/down arrows push buttons¿ illuminated, however the front panel touch screen remained blank.An internal inspection of the cycler found transformer (t1) on the ¿inverter board¿ to have internal short.The inverter board is located on the rear of the touch screen.A known good inverter board was installed and the touch screen became operational.The functioning inverter board was removed from the touch screen at the completion of the investigation.There were no other discrepancies encountered in the internal inspection of the cycler.A review of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements.Upon completion of the evaluation, the reported issue was confirmed and the cause was determined to be a burnt transformer on the inverter board.The cycler was refurbished following the evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBERTY SELECT CYCLER ASSY(NON-VALUATED)
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
matthew amaral
920 winter st.
waltham, MA 02451
7816999758
MDR Report Key8645794
MDR Text Key146230328
Report Number2937457-2019-01682
Device Sequence Number1
Product Code FKX
UDI-Device Identifier00840861102068
UDI-Public00840861102068
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181108
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup,Followup
Report Date 06/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/28/2019
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberRTLR180343
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2019
Is the Reporter a Health Professional? Yes
Device Age MO
Date Manufacturer Received06/03/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/12/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-