CONCORD MANUFACTURING LIBERTY SELECT CYCLER ASSY(NON-VALUATED); SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
|
Back to Search Results |
|
Catalog Number RTLR180343 |
Device Problem
Thermal Decomposition of Device (1071)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 04/08/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Plant investigation: the actual device was returned to the manufacturer for physical evaluation.A visual inspection of the returned cycler exterior showed no sign of physical damage.The touch screen test failed - when powering on the cycler the ¿ok, stop, and up/down arrows push buttons¿ illuminated, however the front panel touch screen remained blank.An internal inspection of the cycler found transformer (t1) on the ¿inverter board¿ to have internal short.The inverter board is located on the rear of the touch screen.A known good inverter board was installed and the touch screen became operational.The functioning inverter board was removed from the touch screen at the completion of the investigation.There were no other discrepancies encountered in the internal inspection of the cycler.A review of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformance during the manufacturing process.In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements.Upon completion of the evaluation, the reported issue was confirmed and the cause was determined to be a shorted transformer on the inverter board.The cycler was refurbished following the evaluation.
|
|
Event Description
|
It was reported that the screen of a patient¿s liberty select cycler went blank prior to their dialysis (pd) treatment.At that point in time, the technical support representative advised the patient to discontinue use of the cycler and to notify their peritoneal dialysis registered nurse (pdrn) of the event.A replacement cycler was issued to the patient.It was reported that an alternate treatment option was available.The cycler was returned to the manufacturer and a replacement cycler was provided.Upon physical evaluation of the cycler by the manufacturer, it was identified that the transformer on the inverter board was shorted.
|
|
Event Description
|
It was reported that the screen of a patient¿s liberty select cycler went blank prior to their dialysis (pd) treatment.At that point in time, the technical support representative advised the peritoneal dialysis nurse (pdrn) that was assisting the patient to discontinue use of the cycler.A replacement cycler was issued to the patient.It was reported that an alternate treatment option was available.The cycler was returned to the manufacturer and a replacement cycler was provided.Upon physical evaluation of the cycler by the manufacturer, it was identified that the transformer on the inverter board was shorted.
|
|
Manufacturer Narrative
|
Corrected plant investigation: the actual device was returned to the manufacturer for physical evaluation.A visual inspection of the returned cycler exterior showed no sign of physical damage.The touch screen test failed - when powering on the cycler the ¿ok, stop, and up/down arrows push buttons¿ illuminated, however the front panel touch screen remained blank.An internal inspection of the cycler found transformer (t1) on the ¿inverter board¿ to have internal short.The inverter board is located on the rear of the touch screen.A known good inverter board was installed and the touch screen became operational.The functioning inverter board was removed from the touch screen at the completion of the investigation.There were no other discrepancies encountered in the internal inspection of the cycler.A review of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements.Upon completion of the evaluation, the reported issue was confirmed and the cause was determined to be a shorted transformer on the inverter board.The cycler was refurbished following the evaluation.
|
|
Manufacturer Narrative
|
Removed consumer from report source plant investigation: the actual device was returned to the manufacturer for physical evaluation.A visual inspection of the returned cycler exterior showed no sign of physical damage.The touch screen test failed - when powering on the cycler the ¿ok, stop, and up/down arrows push buttons¿ illuminated, however the front panel touch screen remained blank.An internal inspection of the cycler found transformer (t1) on the ¿inverter board¿ to have internal short.The inverter board is located on the rear of the touch screen.A known good inverter board was installed and the touch screen became operational.The functioning inverter board was removed from the touch screen at the completion of the investigation.There were no other discrepancies encountered in the internal inspection of the cycler.A review of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements.Upon completion of the evaluation, the reported issue was confirmed and the cause was determined to be a burnt transformer on the inverter board.The cycler was refurbished following the evaluation.
|
|