Model Number 5392
Device Problems Corroded (1131); Calibration Problem (2890)
Patient Problem No Patient Involvement (2645)
Event Date 05/13/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the rate knob of the external pulse generator (epg) appeared corroded, and it was stuck.It was suggested that the epg be returned for service.There was no patient involvement.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product analysis : analysis confirmed customer comment that pacing rate knob is stuck and corroded.Keypad and one knob are corroded.Lower case is broken.One case screw is contaminated.All found defective parts were replaced and all other identified issues were resolved.Passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUAL CHAMBER TEMPORARY PACEMAKER
Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key8649324
MDR Text Key146343566
Report Number3004593495-2019-00498
Device Sequence Number1
Product Code DTE
UDI-Device Identifier00763000162443
UDI-Public00763000162443
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 10/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/29/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Yes
Date Manufacturer Received10/09/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse