• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS SAFESTEP HUBER NEEDLE SET 22G X 0.75IN; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS SAFESTEP HUBER NEEDLE SET 22G X 0.75IN; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number N/A
Device Problems Mechanical Problem (1384); Fail-Safe Did Not Operate (4046)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/03/2019
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned, at this time, to the manufacturer for evaluation.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.Device not yet returned.
 
Event Description
It was reported that the base did not move down and the safety mechanism didn¿t work.There was no reported patient injury.
 
Event Description
It was reported that the base did not move down and the safety mechanism didn¿t work.There was no reported patient injury.
 
Manufacturer Narrative
H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of safety mechanism failure to activate was confirmed and the cause was determined to be supplier related.The product returned for evaluation was one 22ga x 0.75" safestep safety infusion set.The investigation findings were consistent with misplaced or excess manufacturing adhesive, which bound the safety mechanism to the needle.The following observations were made during the sample evaluation: the safety mechanism was not engaged over the needle tip and was returned positioned at the needle base.The needle shaft and safety mechanism were both examined and were determined to be undamaged and thus were not suspected to have contributed towards the event.Microscopic examination under uv light assistance revealed a white/clear adhesive-like substance at the interface of the needle shaft and safety mechanism.That material fluoresced under ultraviolet light, which was consistent with the adhesive used to assemble the infusion set.From this, it appeared that adhesive was deposited on the needle shaft during device manufacture.The device is a supplied component and the supplier was notified of the event.Corrective actions have been implemented by the supplier; however, it is unknown if the complaint sample was manufactured prior to or following that implementation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFESTEP HUBER NEEDLE SET 22G X 0.75IN
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8649561
MDR Text Key146631351
Report Number3006260740-2019-01489
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00801741066160
UDI-Public(01)00801741066160
Combination Product (y/n)N
PMA/PMN Number
K153440
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 06/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/29/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberLH-0029
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/29/2019
Event Location Hospital
Date Manufacturer Received06/07/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-