Model Number 2420-0007
Device Problem Fluid/Blood Leak (1250)
Patient Problems Venipuncture (2129); No Consequences Or Impact To Patient (2199)
Event Date 05/06/2019
Event Type  malfunction  
Manufacturer Narrative
The devices have been received and the evaluation is pending.A follow up report will be submitted with investigation results once the evaluation has been completed.
 
Event Description
The reported feedback suggests that there are connection issues.From the reported information there are no indications of serious injury to the patient or user as a result of this incident.
 
Manufacturer Narrative
The customer's report of connection issues was confirmed.A visual inspection confirmed the customer's report as the sample was received without a male luer.A closer inspection did not identify any solvent around the end of the tubing.The root cause of the customer's report was due to an insufficient amount of solvent having been applied to the tubing during the assembly process.
 
Event Description
The reported feedback suggests that there are connection issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS® PUMP MODULE ADMINISTRATION SET
Type of Device
SET,ADMINISTRATION,INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
MDR Report Key8655953
MDR Text Key146632249
Report Number9616066-2019-01503
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
PMA/PMN Number
K944320
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 05/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/30/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2420-0007
Device Catalogue Number2420-0007
Device Lot Number18115925
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/21/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-