Other relevant device(s) are: product id: 9733571, serial/lot #: (b)(4).
A medtronic representative went to the site to test the equipment.
Testing revealed that the system performed as intended.
The system then passed the system checkout and was found to be fully functional.
The cable was returned to the manufacturer for analysis.
Analysis found that a continuity test revealed an open from pin 15 of the db26 connector to pin 6 of the fischer connector.
Analysis found that the reported event was related to an electrical issue.
The hardware investigation found that the reported event was related to a hardware issue.
This issue was documented in a medtronic navigation hardware anomaly tracking database.
If information is provided in the future, a supplemental report will be issued.
|