• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC STEALTHSTATION S7 SYSTEM; INSTRUMENT, STEREOTAXIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC STEALTHSTATION S7 SYSTEM; INSTRUMENT, STEREOTAXIC Back to Search Results
Model Number 9734056
Device Problems Electrical /Electronic Property Problem (1198); Audible Prompt/Feedback Problem (4020)
Patient Problem No Patient Involvement (2645)
Event Date 04/19/2019
Event Type  malfunction  
Manufacturer Narrative
Other relevant device(s) are: product id: 9733571, serial/lot #: (b)(4).A medtronic representative went to the site to test the equipment.Testing revealed that the system performed as intended.The system then passed the system checkout and was found to be fully functional.The cable was returned to the manufacturer for analysis.Analysis found that a continuity test revealed an open from pin 15 of the db26 connector to pin 6 of the fischer connector.Analysis found that the reported event was related to an electrical issue.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation device being used outside a procedure.It was reported that there was a loud noise from the speaker of the touch screen monitor.There was no patient present when this issue was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
INSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
stacy ruemping
7000 central avenue ne rcw215
minneapolis, MN 55432
7635260594
MDR Report Key8656333
MDR Text Key146644028
Report Number1723170-2019-02801
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/30/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date09/26/2016
Device Model Number9734056
Device Catalogue Number9734056
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/30/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/08/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-