Intra-aortic balloon (iab) therapy was started in a patient with cardiopulmonary arrest.Condensation was observed inside of the catheter tubing and was manually removed.15 hours later, blood was observed in the tubing.The console generated an iab catheter restriction alarm.The iab catheter was attempted to be retracted in the sheath but it could not be pulled back into the sheath.Thus both the iab catheter and the sheath were required to remove as a unit from the patient.Prior to removal of the devices, a guidewire was attempted to be delivered through the iab catheter in order to remain the access to the proper position for placing in the descending aorta for the replacement device.However, the guidewire got stuck within the iab catheter during its delivery.Therefore, whole device including the guidewire, iab catheter and the sheath was removed from the patient.After removal of the devices, when the retrieved iab catheter was checked visually outside of the patient, the inner lumen of the balloon was found to have been damaged around its proximal portion.Another iab catheter was used instead to continue the iabp therapy.During therapy, the physician determined that the patient would not be able to recover and therapy was discontinued with the consent of the family.Approximately one to two days after discontinuation of the iabp therapy, the patient was reported to expire.The facility does not attribute the death to either device.This report is for the second iab used with no reported malfunction.
|