• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONCORD MANUFACTURING 2008K2 HEMODIALYSIS SYS OLC/DP, SPANISH; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING 2008K2 HEMODIALYSIS SYS OLC/DP, SPANISH; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 190618
Device Problem Melted (1385)
Patient Problem No Patient Involvement (2645)
Event Date 05/17/2019
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event Description
A customer reported to fresenius (b)(4) that a 2008k2 machine was unable to turn on.There was no patient involvement since the event occurred before treatment and the patient was not connected.A fresenius (b)(4) technician was scheduled to service the reported machine.A fresenius (b)(4) technician replaced the melted fuses and conducted functionary tests.The machine was left function as intended.There were no parts being returned.
 
Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation.A fresenius mexico technician performed an on-site inspection and found the fuses had sustained heat damage.A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation into the cause of the reported problem was not able to be confirmed.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008K2 HEMODIALYSIS SYS OLC/DP, SPANISH
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
MDR Report Key8662280
MDR Text Key146794429
Report Number2937457-2019-01753
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
PMA/PMN Number
K153449
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 06/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/03/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number190618
Was Device Available for Evaluation? No
Device Age MO
Date Manufacturer Received06/14/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-