Catalog Number R5C4479C
Device Problems Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/08/2019
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the patient line of a homechoice cassette leaked; further described as ¿it was squirting out of the hole." this was observed during drain one of five of peritoneal dialysis therapy; the patient was connected.Renal therapy services (rts) assisted the patient in ending therapy and advised to start over with all new supplies.Rts advised to inform their peritoneal dialysis registered nurse (pdrn) regarding the event.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information: two actual (2) samples were received for evaluation.A visual inspection was performed with the naked eye noted a cuts in the tubing on both samples.One (1) sample had cuts in the tubing approximately 22 inches from the blue shrouder on the heater line.The other sample had a cut approximately 37 inches from the patient connector on the patient line.Functional testing including clear passage and clamp function testing were performed with no issues noted.Leak testing revealed a leak through the cuts of both samples.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
mountain home AR
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key8663923
MDR Text Key146854186
Report Number1416980-2019-03039
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/03/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/04/2024
Device Catalogue NumberR5C4479C
Device Lot NumberH19B04072
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2019
Is the Reporter a Health Professional? No
Date Manufacturer Received07/26/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
HOMECHOICE
Patient Age65 YR
-
-