AV-TEMECULA-CT MITRACLIP SYSTEM STEERABLE GUIDE CATHETER Back to Search Results
Catalog Number SGC0302
Device Problems Deformation Due to Compressive Stress (2889); Difficult to Advance (2920); Material Split, Cut or Torn (4008)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/15/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4). During processing of this complaint, attempts were made to obtain complete event, patient and device information. The device is expected to be returned for evaluation. It has not yet been received. A follow up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report torn material. It was reported that this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 3-4. The steerable guide catheter (sgc 80925u240) was inserted into the groin with the dilator; however, the sgc could not advance over the wire into the femoral vein. The sgc was retracted when the dilator was pushed back into the sgc soft tip. The shaft became kinked, and the soft tip had an incision-like damage, and appeared to be torn. It was later noted that during preparation of the sgc, force was applied to advance the dilator out of the sgc. The sgc was removed and replaced. A new sgc was used to complete the procedure. One clip was implanted, reducing mr to 1. There were no adverse patient effects and no clinically significant delay in the procedure. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMITRACLIP SYSTEM STEERABLE GUIDE CATHETER
Type of DeviceSTEERABLE GUIDE CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
MENLO PARK, CA REG# 3005070406
abbott vascular
26531 ynez rd.
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key8664220
MDR Text Key147166832
Report Number2024168-2019-04330
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Remedial Action Other
Type of Report Initial,Followup
Report Date 07/09/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received06/03/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date09/26/2019
Device Catalogue NumberSGC0302
Device Lot Number80925U240
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received06/26/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-