Model Number AB-5100R |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Complaint, Ill-Defined (2331)
|
Event Date 05/07/2019 |
Event Type
malfunction
|
Event Description
|
The recipient's device was explanted.The recipient was reimplanted with another cochlear device.Advanced bionics is in the process of obtaining additional information.When additional information is received, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
The external visual inspection revealed the electrode was cut prior to receipt.This is believed to have occurred during revision surgery.The photographic imaging inspection confirmed cut electrode wires.This is believed to have occurred during revision surgery.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The recipient reportedly elected revision surgery for a technology upgrade.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|