It was reported that the user had noted several issues while using the mobile programmer and its affiliated application.The battery on the user's tablet was draining too fast, about 1% per minute and it was speculated that it could be due to the variable gain amplifier (vga) slave cable.It was additionally noted that when plugged in the tablet does not charge and at a battery level of 23% the tablet tells the user it must be plugged in and continues to generate the error message.It was also noted that even in airplane mode it was not charging sufficiently.It was further reported that clearing analyzer measurements crashes the application and that there are connectivity issues with the implantable heart device, that it loses connectivity several times per [implant procedure] case and that the user experienced one incident of being unable to re-establish the connection without a reset.Finally it was reported that reports generate slowly.The mobile programmer and the application remain in use.No patient complications have been reported as a result of this event.
|