• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINN STRAIGHT CUP IMPACTOR; HIP INSTRUMENTS : IMPACTORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINN STRAIGHT CUP IMPACTOR; HIP INSTRUMENTS : IMPACTORS Back to Search Results
Catalog Number 221750041
Device Problems Break (1069); Material Deformation (2976); Scratched Material (3020); Unintended Movement (3026)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/17/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the strike site of the insertion handle was broke during impaction of the implant.After the first three strikes with the malet, the weld connecting the strike site of the but end of the insertion handle broke.This ultimately resulted in the strike site breaking off from the body of the insertion handle.No surgical delay.
 
Manufacturer Narrative
B)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.  investigation summary: examination of the returned instrument found the impactor damaged.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history review
=
> null if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN STRAIGHT CUP IMPACTOR
Type of Device
HIP INSTRUMENTS : IMPACTORS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key8671894
MDR Text Key147097320
Report Number1818910-2019-95315
Device Sequence Number1
Product Code MRA
UDI-Device Identifier10603295098980
UDI-Public10603295098980
Combination Product (y/n)N
PMA/PMN Number
P070026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 05/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/05/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number221750041
Device Lot NumberNT0109
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2019
Date Manufacturer Received07/12/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-