• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONCORD MANUFACTURING LIBERTY SELECT CYCLER ASSY(NON-VALUATED); SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING LIBERTY SELECT CYCLER ASSY(NON-VALUATED); SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number RTLR180343
Device Problem Sparking (2595)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/26/2019
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event Description
It was reported that the patient¿s liberty select cycler turns on and off on its own throughout their peritoneal dialysis (pd) treatment.The power cord was properly connected in both ends and there were no recent power outages in the home or in the area reported.The patient reported seeing sparks coming from the power cord.The patient has been able to complete treatment.At that point in time, the technical support representative advised the patient to discontinue use of the cycler and to notify their peritoneal dialysis registered nurse (pdrn) of the event.A replacement cycler was issued to the patient.It was reported that an alternate treatment option was available.Upon follow up, the patient stated that she saw a flash/flashing coming from the power cord.The patient confirmed that there were no adverse events or medical intervention required as a result of the reported event.The patient stated there was no damage to the power outlet or the cycler and that the power plug was not burned.There was no burning smell, melting, smoke, or flame.The patient was able to complete peritoneal dialysis treatment using continuous ambulatory peritoneal dialysis (capd) in the absence of the cycler.The patient has received a new cycler which is working well and is continuing peritoneal dialysis therapy with no further issues.The cycler was reported to be returned to the manufacturer.
 
Manufacturer Narrative
Plant investigation: the actual device was returned to the manufacturer for physical evaluation.An external visual inspection was performed on the cycler with no physical damage noted.The reported symptoms of ¿sparks and cycler turns on/off¿ was not confirmed.The voltage verification check passed.The simulated treatment pre-accelerated stress test (ast) 15 minute 1000 ml was performed and completed without any failures or problems.An internal visual inspection of the returned cycler revealed no other discrepancies.A review of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements.Upon completion of the evaluation, there were no malfunctions that could have caused or contributed to the reported event.The cycler performed as designed and an associated cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBERTY SELECT CYCLER ASSY(NON-VALUATED)
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
MDR Report Key8678177
MDR Text Key147314025
Report Number2937457-2019-01807
Device Sequence Number1
Product Code FKX
UDI-Device Identifier00840861102068
UDI-Public00840861102068
Combination Product (y/n)N
PMA/PMN Number
K181108
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 06/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/07/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberRTLR180343
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/03/2019
Device Age MO
Date Manufacturer Received06/10/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
DELFLEX PD FLUID; LIBERTY CYCLER SET; DELFLEX PD FLUID; LIBERTY CYCLER SET
Patient Age47 YR
Patient Weight83
-
-