The investigation determined that non-reproducible vitros phyt (phenytoin) results were obtained from two different samples collected from the same patient tested on two different vitros 5600 integrated systems.The investigation was unable to determine a definitive assignable cause for this event, however, a sample matrix issue related to the affected patient cannot be ruled out.The issue is isolated only to samples collected from the affected patient, as other patient samples run over this time frame did not demonstrate a similar issue.Quality control and precision test results from the time frame of the event demonstrated acceptable within-laboratory and within-run performance of the vitros phyt slide lot 2618-0169-6447.Ongoing tracking and trending of complaints has not identified any signals that would suggest there is a systemic issue with vitros phyt lot 2618-0169-6447.With the exception of the non-reproducible results obtained from the affected patient samples, vitros phyt lot 2618-0169-6447 otherwise appeared to be performing within expectations.In addition, results of precision testing demonstrated that one of the vitros 5600 integrated systems was operating as expected at the time of the event.Finally, it is unknown if the affected sample was centrifuged according to the collection device manufacture¿s recommendations, therefore, improper pre-analytical sample handling cannot be ruled out as a contributing factor.It is possible that cellular debris, due to poor sample preparation, was present in the affected samples, although this could not be confirmed.
|
A customer obtained non-reproducible vitros phyt (phenytoin) results from two different samples collected from the same patient tested on two different vitros 5600 integrated systems.On (b)(6) 2019: patient sample results of greater than 158.4, 200.0, 207.0, greater than 158.4, and greater than 158.4 umol/l vs.The expected result of 124.5 umol/l.On (b)(6) 2019: patient sample result of 97 umol/l vs.The expected result of 172.0 umol/l biased results of the direction and magnitude observed may lead to inappropriate physician action if undetected.The non-reproducible vitros phyt patient sample results were not reported from the laboratory.Ortho has not been made aware of any allegation of actual patient harm as a result of this event.This report corresponds to ortho clinical diagnostics inc (ortho).Complaint numbers (b)(4).
|