• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIAGNOSTICS DIVISION CELL-DYN RUBY ANALYZER; AUTOMATED HEMATOLOGY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIAGNOSTICS DIVISION CELL-DYN RUBY ANALYZER; AUTOMATED HEMATOLOGY ANALYZER Back to Search Results
Catalog Number 08H67-01
Device Problem High Readings (2459)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
All available patient information has been included.No additional patient information is available.An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event Description
The customer reported discrepant hemoglobin results for a male patient when processing on the cell-dyn ruby.The customer reported the patient has a known history of low hemoglobin and platelet results.The initial hemoglobin result was 4.79 and retest using the same analyzer was 9.41.The customer repeated the sample on a different ruby analyzer and the hemoglobin result was 4.29.The customer uses a normal range of 14.0-18.0 for hemoglobin.No impact to patient management was reported.
 
Manufacturer Narrative
An investigation was performed for the customer issue and included a review of the complaint text, review of product historical data, troubleshooting, and review of product labeling.Review of product historical data for any trends and all customer complaints received for this issue did not identify any adverse trends or abnormal complaint activity.Based on complaint information, the field service engineer replaced the following parts: a.8921204903, shear valve assembly b.8921016301, hgb lyse - v28 nc valve c.9130543, hgb pathway tubing after the troubleshooting, the instrument passed precision test and control were within range.Labeling was reviewed and found to be adequate.Based on all available information and abbott diagnostics' complaint investigation a product deficiency was not identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL-DYN RUBY ANALYZER
Type of Device
AUTOMATED HEMATOLOGY ANALYZER
Manufacturer (Section D)
ABBOTT DIAGNOSTICS DIVISION
4551 great america parkway
santa clara CA 95054
MDR Report Key8679104
MDR Text Key169927310
Report Number2919069-2019-00066
Device Sequence Number1
Product Code GKZ
UDI-Device Identifier00380740017170
UDI-Public00380740017170
Combination Product (y/n)N
PMA/PMN Number
K061667
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 07/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/07/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number08H67-01
Was Device Available for Evaluation? Yes
Date Manufacturer Received07/22/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-