Model Number SN6CWS |
Device Problem
Particulates (1451)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/21/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
A sample product was not received for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There are no other complaints in the lot number.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A surgeon reported that during an intraocular lens (iol) implant procedure, there was white foreign things in the lens.Additional information was requested.
|
|
Manufacturer Narrative
|
A photo was returned for analysis.The photo provided shows the lens in the eye.The photo has an area circled in yellow showing what is interpreted as the foreign material.The area has the appearance of a scratch.Without the actual sample no determination can be confirmed.Based on our observation of the attached photos, the lens appears to have damage on the optic surface.It is difficult to make a final determination without evaluation of the physical sample.It is unknown if the qualified viscoelastic was used.The root cause cannot be determined based on available information.The manufacturer internal reference number is: (b)(4).
|
|