• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL W/ACRYSERT C DELIVERY SYSTEM; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL W/ACRYSERT C DELIVERY SYSTEM; INTRAOCULAR LENS Back to Search Results
Model Number SN6CWS
Device Problem Particulates (1451)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/21/2019
Event Type  malfunction  
Manufacturer Narrative
A sample product was not received for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There are no other complaints in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event Description
A surgeon reported that during an intraocular lens (iol) implant procedure, there was white foreign things in the lens.Additional information was requested.
 
Manufacturer Narrative
A photo was returned for analysis.The photo provided shows the lens in the eye.The photo has an area circled in yellow showing what is interpreted as the foreign material.The area has the appearance of a scratch.Without the actual sample no determination can be confirmed.Based on our observation of the attached photos, the lens appears to have damage on the optic surface.It is difficult to make a final determination without evaluation of the physical sample.It is unknown if the qualified viscoelastic was used.The root cause cannot be determined based on available information.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ SINGLEPIECE IOL W/ACRYSERT C DELIVERY SYSTEM
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
MDR Report Key8681273
MDR Text Key147455850
Report Number1119421-2019-00846
Device Sequence Number1
Product Code HQL
Combination Product (y/n)N
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 08/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/09/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2023
Device Model NumberSN6CWS
Device Catalogue NumberSN6CWS.230
Device Lot Number12645248
Was Device Available for Evaluation? No
Date Manufacturer Received08/08/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-