Model Number MMT-1715KR |
Device Problem
No Display/Image (1183)
|
Patient Problem
Bruise/Contusion (1754)
|
Event Date 06/02/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
|
Event Description
|
Customer reported via phone call that the insulin pump had a blank display.The customer¿s blood glucose level was 110 mg/dl at the time of the incident.The customer was assisted with troubleshooting and customer reports display was blank currently.Customer states the display was not blank less than 30 seconds and did not return.The customer removes the battery and stated the insert battery alarm did not display on the insulin pump screen.Customer states the contact on the battery cap was neither missing nor damage.Customer states battery compartment was neither damaged nor corroded.Customer states the spring was neither damaged nor corroded.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be returned for analysis.
|
|
Manufacturer Narrative
|
Device passed sleep current measurement, active current measurement, self test and displacement test.No unexpected low battery alarm, battery power loss alarm or blank display noted during testing.(b)(4).
|
|