• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD. 8010379 ADAPTER SLEEVES 12/14 +5; ASR HIP SYSTEM : HIP FEMORAL SLEEVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD. 8010379 ADAPTER SLEEVES 12/14 +5; ASR HIP SYSTEM : HIP FEMORAL SLEEVE Back to Search Results
Catalog Number 999800315
Device Problems Corroded (1131); Biocompatibility (2886)
Patient Problems Foreign Body Reaction (1868); Pain (1994); Tissue Damage (2104); Discomfort (2330); Injury (2348); Test Result (2695); No Code Available (3191)
Event Date 10/02/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Occupation: lawyer.
 
Event Description
Asr litigation records alleging injury, metallosis, pain, emotional distress and elevated blood metal ions.Ppd and medical records received.After review of medical records, the patient was revised to address left hip mechanical failure.Operative findings include: trunnionosis with black metal staining to the trunnion and taper of the femoral head, chronic synovitis consistent with metal failure and less than 10 cc of yellow, slightly cloudy fluid.The fluid was sent for culture and frozen section.These were negative for any acute inflammation and cell count was approximately 2700 with standard cell counter and slightly over 200 with manual differential.There was no acute synovitis but had gray, almost granular type of tissue seen with metal failure.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #
=
> (b)(4).Investigation summary
=
> this hip replacement platform was voluntarily recalled from the market and the product codes are now considered inactive.Further investigation of this individual incident will not be undertaken.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADAPTER SLEEVES 12/14 +5
Type of Device
ASR HIP SYSTEM : HIP FEMORAL SLEEVE
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD. 8010379
st. anthony's rd
leeds LS11 8 DT
UK  LS11 8 DT
MDR Report Key8687807
MDR Text Key147622792
Report Number1818910-2019-95716
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
PMA/PMN Number
K040627
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 10/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/11/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number999800315
Device Lot Number2746879
Was Device Available for Evaluation? No
Date Manufacturer Received08/26/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-1749/1816-2011
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age58 YR
-
-