(b)(6).
Returned product consisted of a quantum maverick balloon catheter.
The shaft, hypotube, tip and balloon were microscopically and visually examined.
The balloon was tightly folded.
There were numerous hypotube kinks throughout the device.
There was a complete hypotube separation 48cm distal of the strain relief.
The hypotube fracture surfaces were ovaled, which suggests the device was kinked prior to separation.
Inspection of the remainder of the device presented no other damage or irregularities.
|