Patient allegedly received an implant on (b)(6) 2003 due to recurrent deep vein thrombosis with pulmonary embolism.Patient alleges migration, tilt, vena cava perforation.Patient further alleges stress, anxiety, pain, swollen leg and swollen feet.Per ct, (b)(6) 2017: inferior vena cava filter has migrated cephalad with the proximal tip penetrating 9 mm beyond the lumen and 22 degree rightward tilt relative to the ivc axis coursing medial to the right renal vein and anterior to the right renal artery.A left anterolateral ivc arm penetrates 10 mm beyond the lumen extending into the posterior wall of the duodenum.Negative for ivc filter arm fracture, occlusion, or thrombus.
|
Additional information: investigation ¿ the following allegations have been investigated: migration, tilt, vc/organ perf, stress, anxiety, pain, swollen legs, swollen feet.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Vena cava wall penetration/perforation has been reported and may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation.Filter or filter fragment migration and (or) embolization (e.G., movement to the heart or lungs) has been reported.Filter or filter fragment movement has occurred in both the cranial and caudal direction and may be either symptomatic or asymptomatic.Potential causes may include filter placement in ivcs with diameters smaller or larger than those specified in these instructions for use; improper deployment; deployment into thrombus; dislodgement due to large thrombus burdens; and (or) excessive force or manipulations near an in situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: filter migration, trauma to adjacent structures.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported stress, anxiety, pain, swollen feet, swollen legs are directly related to the filter and unable to identify a corresponding failure mode at this point in time.Catalog number and lot number are unknown; however, there is no evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.Blank fields on this form indicate the information is unknown or unavailable, or unchanged.
|