The device was not returned to the service center for evaluation.
The user facility did not provide a specific serial number for the subject; therefore, no review of the instrument history can be performed.
Olympus will continue to investigate this complaint to obtain more detailed information regarding the reported events.
If additional information becomes available, this report will be updated and supplemented accordingly.
The cyf-vh ¿reprocessing before the first use/reprocessing and storage after use¿ section of the instruction manual warns users after using this instrument, reprocess and store it according to the instructions given in the endoscope¿s companion reprocessing manual.
Improper and/or incomplete reprocessing or storage can pose an infection control risk, cause equipment damage, or reduce performance.
¿.
|
The service was informed that six patients contracted salmonella infections in their bladder after undergoing cystoscopy procedures with the facility¿s cysto-nephro videoscope.
The user facility¿s physician reported that this has led to sepsis in some of the patients.
The bacteria has been tested and found to be the same genetic bug in each case.
The user facility has identified 35 additional patients that will need to be notified as they also were in contact the subject scope.
All patients were seen at the same clinic.
In addition, the physician reported that the scope passed the leak test and appeared in satisfactory condition upon quick visual inspection.
The facility performed a closer visual inspection on the subject scope and observed break, separation or perforation along the insertion tube of the scope; where the uppermost point of the bending rubber begins and meets the bending section bending rubber.
The scope was then cultured and tested positive for salmonella.
This is 2 of 6 reports.
|