ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER LAUREATE WORLD PHACO SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065750502
Device Problems Intermittent Continuity (1121); Unexpected Shutdown (4019)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/03/2019
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A customer reported that before a cataract procedure, the system intermittently shut down.
 
Manufacturer Narrative
The company service representative examined the system and was not able to confirm or replicate the reported event.As a preventative measure, the power supply was replaced.The system was then tested and met all product specifications.The system manufacturing device history record (dhr) was reviewed.Based on qa assessment, the product met specifications at the time of release.The root cause of the reported event cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAUREATE WORLD PHACO SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key8707545
MDR Text Key148635826
Report Number2028159-2019-01113
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K021566
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/18/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065750502
Was Device Available for Evaluation? Yes
Date Manufacturer Received08/06/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-