Model Number 45031
Device Problems Crack (1135); Display or Visual Feedback Problem (1184); Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/29/2019
Event Type  malfunction  
Event Description
It was reported there was a crack in the catheter shaft.An angiojet solent omni catheter was selected for a thrombectomy procedure of the lower leg.During the procedure, the catheter was inserted and power pulse was completed.A half hour later when thrombectomy was being done, an error message occurred.The catheter was pulled out of the patient and it was noted that there was a crack with a leak in the mid-shaft of the catheter.The procedure was completed with another of the same catheter.There were no patient complications and the patient's status post procedure was fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOJET SOLENT OMNI
Type of Device
CATHETER, EMBOLECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC SCIMED, INC
two scimed place
maple grove MN 55311
Manufacturer Contact
sonali arangil
two scimed place
maple grove, MN 55311
6515827403
MDR Report Key8708602
MDR Text Key148322699
Report Number2134265-2019-07030
Device Sequence Number1
Product Code KRA
UDI-Device Identifier08714729889663
UDI-Public08714729889663
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111182
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 06/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/18/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/16/2020
Device Model Number45031
Device Catalogue Number45031
Device Lot Number0022477235
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/29/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/18/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-