Livanova (b)(4) manufactures the s5 control panel mrp 85.The incident occurred in (b)(6).During preventive maintenance, perfusionist requested certain parts be replaced as preemptive measure due to a previous unreported problem.A livanova field service representative was informed of the issue during the preventive maintenance for this unit.During preventive maintenance, perfusionist requested certain parts be replaced as preemptive measure due to a previous unreported problem.Livanova field service representative replaced the processor board, the ribbon and the flat cables, the led display and the cable speaker.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|