PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 20 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 20
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/31/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).The device was not returned to physio-control for an evaluation.The cause of the reported issue could not be determined.Device not evaluated by manufacturer.
 
Event Description
The customer contacted physio-control to report that during a synchronized cardioversion procedure, the patient had a very complicated rhythm.The sync button was engaged but the device discharged on the "t" wave causing the patient to go into ventricular fibrillation.They were able to change the device to lead 2 and convert the patient out of that rhythm.There were no reports of adverse effects to the patient.Physio-control has made multiple attempts to contact the customer in order to obtain additional information about both the patient and the event; however, no response has been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 20 DEFIBRILLATOR/MONITOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key8714592
MDR Text Key148515701
Report Number0003015876-2019-01058
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130454
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 06/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/19/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number20
Device Catalogue NumberUNK_SMP
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/31/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-