• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY CMW - 9610921 SMARTSET GMV 40G US EO; BONE CEMENT : BONE CEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY CMW - 9610921 SMARTSET GMV 40G US EO; BONE CEMENT : BONE CEMENT Back to Search Results
Model Number 5450-50-501
Device Problem Loss of or Failure to Bond (1068)
Patient Problems Adhesion(s) (1695); Edema (1820); Muscular Rigidity (1968); Pain (1994); Loss of Range of Motion (2032); Scarring (2061); Swelling (2091); Synovitis (2094); Limited Mobility Of The Implanted Joint (2671); No Code Available (3191)
Event Date 08/01/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Medical records received 23 jul 2018.Medical records were reviewed for mdr reportability.Patient underwent a right tka with depuy cement.She then underwent a closed manipulation for decreased range of motion.Patient was then revised for ongoing pain and swelling.Intraoperative findings included synovitis, scarring, loose tibial component at implant-cement interface, tibial subsidence medially, a small fracture along the anterior tibia with no indication of cause, and a small proximal tibia metaphyseal cyst.Doi: (b)(6) 2015.Dor: (b)(6) 2017.
 
Manufacturer Narrative
Product complaint # (b)(4).(b)(4).Is used to capture the surgical intervention and medical device removal.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Medical record ad 17 june 2019 was reviewed on 6 december 2019.(b)(6) 2015: the patient underwent a right total knee arthroplasty secondary to severe degenerative joint disease, valgus deformity and tricompartmental changes.Attune implants were used with depuy cement x2.The patella was resurfaced.No intraoperative complications were noted.(b)(6) 2017: the patient underwent a revision of the right knee secondary to pain, stiffness, swelling and suspected tibial loosening.Intraoperatively, the surgeon found synovitis; scar tissue; gross tibial loosening at the cement to implant interface; a small crack in the anterior tibia with no intervention noted; and a small anteromedial proximal tibial metaphyseal cyst.Although the femoral component was well fixed, the surgeon opted to remove it anyway.There were no intraoperative complications.Pmh: atrial fibrillation, pacemaker, sleep apnea, severe degenerative joint disease (right knee).Doi: (b)(6) 2015.Dor: (b)(6) 2017; (right knee).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #
=
> (b)(4).Investigation summary
=
> no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: (b)(6) 2021: the investigation was re-opened upon receipt of additional information.No device associated with this report was received for examination.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot: a device history record (dhr) review was conducted previously on legacy complaint (b)(4).The dhr review found no non-conformance on this batch.Final micro and sterility tests passed.Based on the provided dhr review there were no related manufacturing deviations or anomalies on the given product and lot combination.¿dmf# - (b)(4).¿trade name ¿ (b)(4).¿active ingredient(s) ¿ gentamicin sulphate.¿dosage form - powder.¿strength ¿ 1.0g active in our cements.H10 additional narrative: added: d7a, d9 corrected: d2b, d3, g1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTSET GMV 40G US EO
Type of Device
BONE CEMENT : BONE CEMENT
Manufacturer (Section D)
DEPUY CMW - 9610921
cornford rd
blackpool IN FY4 4 QQ
UK  FY4 4QQ
MDR Report Key8718786
MDR Text Key148734970
Report Number1818910-2019-96424
Device Sequence Number1
Product Code MBB
UDI-Device Identifier10603295174295
UDI-Public10603295174295
Combination Product (y/n)N
PMA/PMN Number
K081163
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup,Followup,Followup
Report Date 06/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/20/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2016
Device Model Number5450-50-501
Device Catalogue Number545050501
Device Lot Number8010566
Was Device Available for Evaluation? No
Date Manufacturer Received02/08/2021
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age63 YR
-
-