MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number N5C8305C
Device Problem Disconnection (1171)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/20/2019
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the ¿whole entire top cap and connector" of the patient line of a homechoice cassette "came off¿.This occurred during setup of peritoneal dialysis therapy.To resolve the issue the caregiver restarted therapy with all new supplies.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information : one actual device was received for evaluation.A visual inspection with the naked eye was performed which noted an absence of a heat seal bead on the patient line tubing to the connector which caused the patient connector to separate from the tubing.The reported condition was verified.The cause of the condition was determined to be manufacturing related.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NI
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
ni
Manufacturer (Section G)
BAXTER HEALTHCARE CORPORATION
ni
ni
ni
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key8719683
MDR Text Key148972344
Report Number1416980-2019-03376
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 10/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/20/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberN5C8305C
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/11/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/17/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-