• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT20649-1
Device Problem Unexpected Shutdown (4019)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/27/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that a unexpected receiver shut-down occurred.No product or data was provided for evaluation.Confirmation of the allegation and a root cause could not be determined.No injury or medical intervention was reported.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The product was evaluated.The device was externally visually inspected and it passed.The receiver failed to charge and reboot.The receiver did not turn on.The battery was replaced with a known good battery and the receiver turned on.The log was downloaded and reviewed finding an error related to the complaint.The receiver case was opened, the internal inspection failed due to a damaged battery.Battery measurements were taken and they were out of specification due to low voltage.The allegation was confirmed.The root cause was determined to be a damaged battery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
ying chen
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key8720492
MDR Text Key149137970
Report Number3004753838-2019-49666
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/20/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMT20649-1
Device Catalogue NumberSTK-GL-010
Device Lot Number5225150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/05/2019
Is the Reporter a Health Professional? No
Date Manufacturer Received08/14/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/10/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age41 YR
Patient Weight77
-
-