Model Number 74281
Device Problem Unable to Obtain Readings (1516)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/03/2019
Event Type  malfunction  
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event Description
The user facility reported to terumo cardiovascular that during cardiopulmonary bypass, the thermistor did not register temperature.As per the user facility, while they were giving heparin, they discovered that the arterial temperature was not working.The perfusionist replaced the thermistor wire with another and the temperature still did not read.No known impact or consequence to patient.The product was not changed out.Procedure was completed successfully.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on june 21, 2019.  upon further investigation of the reported event, the following information is new and/or changed: d10 (device availability - added date returned to manufacturer) g4 (date received by manufacturer) g7 (indication that this is a follow-up report) h2 (follow-up due to additional information) h3 (device evaluation anticipated by manufacturer - a second follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11.) all available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added exp date).G4 (date received by manufacturer).G7 (indication that this is a follow-up report).H2 (follow-up due to additional information and device evaluation).H3 (device evaluated by manufacturer).H4 (device manufacture date).H6 (identification of evaluation codes 10, 11, 3331, 3259, 4307).Method code #1: 10 - testing of actual/suspected device.Method code #2: 11 - testing of device from same lot/batch retained by manufacturer.Method code #3: 3331 - analysis of production records.Results code: 3259 - improper physical structure.Conclusions code: 4307 - cause traced to component failure.The affected sample was returned for evaluation and was checked to confirm that the arterial temperature port was non-functional.A representative retention sample from the same product/lot number combination was tested and found to be functioning properly.An engineering investigation and capa have been initiated to determined a definitive root cause.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESCRIPTIVE OXY PACK - CAPIOX
Type of Device
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key8724136
MDR Text Key148973346
Report Number1124841-2019-00173
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 07/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/21/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2021
Device Model Number74281
Device Catalogue NumberN/A
Device Lot NumberXC18
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/19/2019
Was the Report Sent to FDA? No
Date Manufacturer Received07/23/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-