Model Number 3716 |
Device Problem
Temperature Problem (3022)
|
Patient Problem
Burning Sensation (2146)
|
Event Date 06/05/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
The results/method and conclusion codes along with investigation results will be provided in the final report.
|
|
Event Description
|
Device 1 of 2.Related manufacturer reference number: 1627487-2019-07264.It was reported that the patient has been experiencing a heating sensation while charging their device.As a result, surgical intervention may take place to address this issue.On (b)(6) 2012 st.Jude medical, neuromodulation division, sent field action letters to patients related to heating while charging and raised awareness of this issue to patients.An increase in prior non-reported heating while charging events and other non-reported events was expected.
|
|
Event Description
|
Device 1 of 2.Related manufacturer reference number: 1627487-2019-07264.It was reported that the patient underwent surgical intervention on (b)(6)2019, wherein the patient's ipg was explanted and replaced.
|
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
Device 1 of 2.Related manufacturer reference number: 1627487-2019-07264.
|
|