ST. JUDE MEDICAL - NEUROMODULATION MOBILE CHARGING SYSTEM; SCS CHARGING SYSTEM Back to Search Results
Model Number 3711
Device Problem Temperature Problem (3022)
Patient Problem Burning Sensation (2146)
Event Date 06/05/2019
Event Type  Injury  
Manufacturer Narrative
This charger model was associated with a field correction.The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
Device 2 of 2.Related manufacturer reference number: 1627487-2019-07261.It was reported that the patient has been experiencing a heating sensation while charging their device.As a result, surgical intervention may take place to address this issue.On (b)(6) 2012 st.Jude medical, neuromodulation division, sent field action letters to patients related to heating while charging and raised awareness of this issue to patients.An increase in prior non-reported heating while charging events and other non-reported events was expected.
 
Event Description
Device 2 of 2.Related manufacturer reference number: 1627487-2019-07261.It was reported that the patient underwent surgical intervention on (b)(6)2019, wherein the patient's ipg was explanted and replaced.
 
Event Description
Device 2 of 2.Related manufacturer reference number: 1627487-2019-07261.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOBILE CHARGING SYSTEM
Type of Device
SCS CHARGING SYSTEM
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key8725526
MDR Text Key148909775
Report Number1627487-2019-07264
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 07/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/22/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/28/2014
Device Model Number3711
Device Lot Number4382915
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/17/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/11/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487/07/26/2012/001-C
Patient Sequence Number1
Treatment
SCS IPG
Patient Outcome(s) Other;
Patient Weight43
-
-