Model Number 60-02-15
Device Problems Loss of Power (1475); Infusion or Flow Problem (2964)
Patient Problem No Patient Involvement (2645)
Event Date 06/14/2018
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(6).The complained unit has been requested by livanova (b)(4) for further investigation, together with the drive unit and the flow sensor.During the evaluation the reported issue could be confirmed.The malfunction was traced to a defective cpu board.The board was replaced to resolve the reported malfunction.Subsequent functional verification testing was completed without further issues and the unit was returned to service.A review of the dhr could not identify any deviations or nonconformities relevant to the issue.
 
Event Description
Livanova (b)(4) received a report that the display of the centrifugal pump system with tubing clamp became blank and the flow stopped during use.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
barbara galasso
14401 w. 65th way
arvada, CO 80004
MDR Report Key8729962
MDR Text Key149095316
Report Number9611109-2019-00474
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K032213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 06/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/25/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-02-15
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/09/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/14/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/15/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-