Model Number CYF-5
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Malfunction  
Manufacturer Narrative

The subject device in this report has not been returned to omsc for evaluation. Omsc reviewed the manufacture history (dhr) of the device and confirmed no irregularity. The exact cause of the reported event could not be conclusively determined at this time. If additional information becomes available, this report will be supplemented.

 
Event Description

Olympus medical systems corp. (omsc) was informed that as a result of routine microbiological testing by the user facility, the sample collected from the subject device tested positive for micrococcus luteus (1cfu/endoscope). The device had been manually reprocessed using peracetic acid. There was no report of infection associated with this report.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameOES CYSTONEPHROFIBERSCOPE
Type of DeviceCYSTONEPHROFIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA 192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key8738815
MDR Text Key149567923
Report Number8010047-2019-02318
Device Sequence Number1
Product Code FAJ
Combination Product (Y/N)N
Reporter Country CodeFR
PMA/PMN NumberK032092
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type HEALTH PROFESSIONAL,USER FACI
Reporter Occupation
Type of Report Initial
Report Date 06/27/2019
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received06/27/2019
Is This An Adverse Event Report? No
Is This A Product Problem Report? No
Device Operator HEALTH PROFESSIONAL
Device MODEL NumberCYF-5
Was Device Available For Evaluation? No
Is The Reporter A Health Professional?
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received06/03/2019
Was Device Evaluated By Manufacturer? No Answer Provided
Date Device Manufactured06/15/2017
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-