• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC UNKNOWN; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC UNKNOWN; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number UNKNOWN
Device Problems Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
Patient Problems Anemia (1706); Internal Organ Perforation (1987); Perforation of Vessels (2135); No Information (3190)
Event Type  Injury  
Manufacturer Narrative
Occupation: non healthcare professional.Investigation ¿ it has not been possible to further investigate or evaluate this alleged event based on the limited information and/or no device failure provided to date.Catalog number and lot number are unknown; no evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.
 
Event Description
It is alleged that the patient received a gunther tulip on (b)(6) 2002.It is alleged that the patient was injured without further explanation.Hospital and medical records have been requested but not yet provided.
 
Event Description
Patient allegedly received an implant on (b)(6) 2002 due to deep vein thrombosis.Patient is alleging organ perforation and anemia.Per report from ct (computed tomography): "ivc filter appears rotated and the proximal tip and several of the distal prongs appears outside the lumen of the ivc.".
 
Manufacturer Narrative
Additional information: investigation: h6 device code(s): appropriate term/code not available (3191) was selected for the alleged device perforation and rotation.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Vena cava wall penetration/perforation has been reported and may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Unknown if the reported anemia and rotation are directly related to the filter and unable to identify a corresponding failure mode at this point in time.The following allegations have been investigated.Vena cava (vc)/organ perforation, anemia and rotation.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key8739526
MDR Text Key149357347
Report Number1820334-2019-01527
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 09/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/27/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received08/12/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Other;
Patient Weight135
-
-