The device was implanted in the patient.Since the device was not returned, we are unable to perform further root cause analysis.All devices are 100% tested and all products are 100% inspected for damages and irregularities during manufacture.No corrective action.Monitoring and trending this type of event.If information is provided in the future, a supplemental report will be issued.
|
Medtronic received report that 2 pipeline flex with shield (ped2) were used to treat a neck recurrence (2 x 3 x 3mm) of a ruptured and coiled pcoma-aneurysm (initially 8x5x4,5 mm).Catheterization of left ica with triaxial system (neuron max, navien 058, phenom 27, transcend platinum).Based on sim and cure calculations a 4 x 10 mm ped was chosen.Unproblematic initial unsheathing of the device in m1 segment (to release the petals) and re-sheathing.Correct positioning of ped2 in the ica terminus to over cover the neck of the aneurysm.Unsheathing of the device.However, re-sheathing of the delivery wire with the microcatheter proved to be difficult (the wire was aligned against the outer curve of cavernous ica) and was aborted.In the control run, there was distal shortening of the ped2 (pli10) was noticed ¿ resulting in insufficient aneurysm neck coverage.A second device of 4 x 12 mm ped (pli20) was delivered in the same manner, with a slightly distal landing zone.In the last run, there was enough neck coverage.
|