A steris service technician arrived onsite following the reported event to inspect the 16" century sterilizer.The technician found that one of the bolts to the s7 valve was broken causing the water to leak and the reported event to occur.The technician replaced the damaged bolt, ran a test cycle, and found the sterilizer to be operating according to specification.The unit was returned to service.The sterilizer was installed in 2007 and is approximately 12 years old.A 2-year complaint review indicates this to be an isolated event.No additional issues have been reported.
|