• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONCORD MANUFACTURING LIBERTY SELECT CYCLER ASSY(NON-VALUATED); SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING LIBERTY SELECT CYCLER ASSY(NON-VALUATED); SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number RTLR180343
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/01/2019
Event Type  malfunction  
Manufacturer Narrative
Plant investigation: the actual device was returned to the manufacturer for physical evaluation.An exterior visual inspection of the returned cycler showed no signs of physical damage.Upon power up, the cycler touch screen test failed.When powering on the cycler, the ok, stop and up/down arrow push buttons illuminated, however the front panel touch screen remained blank.It was identified that the cause for the blank screen was due to an internal short on transformer (t1) of the inverter board.The inverter board is located on the rear of the touch screen.A known good inverter board was installed and the display became fully operational.An internal visual inspection of the returned cycler encountered no other discrepancies.A device history record review found a prior occurrence of dim, distorted, or blank screen.A production floor problem report indicated on (b)(6) 2019 that the failure was due to a defective inverter board.Upon completion of the evaluation, the reported issue was confirmed and the cause was determined to be an internal short on the transformer of the inverter board.The cycler was refurbished following the evaluation.
 
Event Description
It was reported that the screen of a patient¿s liberty select cycler went blank during their peritoneal dialysis (pd) treatment.The power outlet was properly working and the ok and stop keys were on, however the screen remained blank.At that point in time, the technical support representative advised the patient to discontinue use of the cycler and to notify their peritoneal dialysis registered nurse (pdrn) of the event.A replacement cycler was issued to the patient.It was reported that an alternate treatment option was available.Additional information was requested, however to date has not been provided.The cycler was returned to the manufacturer and upon physical evaluation of the cycler, it was identified that there was an internal short on the transformer of the inverter board.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBERTY SELECT CYCLER ASSY(NON-VALUATED)
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
matthew amaral
920 winter st.
waltham, MA 02451
7816999758
MDR Report Key8746866
MDR Text Key149610341
Report Number2937457-2019-02037
Device Sequence Number1
Product Code FKX
UDI-Device Identifier00840861102068
UDI-Public00840861102068
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181108
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 06/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/30/2019
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberRTLR180343
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/07/2019
Device Age MO
Date Manufacturer Received06/24/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/02/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-