According to the reporter, in a neck dissection procedure, during cleaning off the blood of the tip at the end of the procedure, it was noted that the tip was chipped a lot.The fragment have not been found.The part fell into the patient's cavity and could not be retrieved.The procedure was completed with another device.X-ray was used to verify that there were no parts left in the patient's cavity.
|
One device was received for evaluation.The returned product did not meet specification as received because of the broken tip of the jaw.Visual inspection found that the plastic tip on the jaw showed mechanical damage and fractured.The broken pieces were not returned.The reported condition of component disengagement was confirmed.The rfid logs taken from the device indicate the product was used on an ft10 generator.The insulation tip was broken by damage from an external force, consistent with the user inadvertently grasping onto a hard object, where led to the chipping of the insulation.Manufacturing and supplier were ruled out as the possible cause of the fracture by the engineering.The device's mechanical function, jaw gap, and closed circuit resistance were all evaluated and found to be within specification.The device was plugged into an ft10 generator and was recognized.The device was activated ten times on a saline soaked towel with satisfactory results.The investigation identified the root cause of the reported event to be user.This unit does not meet the requirements for a manufacturing failure therefore a device history review is not required.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|