WILLIAM COOK EUROPE GUNTHER TULIP NAVALIGN FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
|
Back to Search Results |
|
Catalog Number IGTCFS-65-1-FEM-TULIP |
Device Problems
Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
|
Patient Problems
Pain (1994); Perforation of Vessels (2135); Disability (2371); No Information (3190)
|
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Catalog# is unknown but referred to as cook gunther tulip filter.Occupation: non-healthcare professional.It has not been possible to investigate or evaluate this alleged event based on the limited information provided to date.Cook will reopen its investigation if further information is receiving warranting supplementation in accordance with 21 c.F.R.803.56.
|
|
Event Description
|
'it is alleged that "[pt] received a cook gunther tulip filter on (b)(6) 2011".It is alleged that [pt] was injured without further explanation.Hospital and medical records have been requested but not yet provided.
|
|
Event Description
|
Patient allegedly received an implant on (b)(6) 2011 due to pulmonary embolism (pe).Patient is alleging vena cava perforation.The patient further alleges ¿significant mental and physical pain and suffering, severe and permanent injuries requiring past and future medical treatment, and resulting in disability, impairment, loss of enjoyment of life, loss of car, comfort, and incurred financial or economic loss, including, but not limited to, obligations for medical expenses, lost income, and other damages.¿ per the post procedure progress note dated (b)(6) 2011: ¿procedure: ivc filter placement (cook tulip)¿; post procedure diagnosis: ¿pulmonary embolism¿; complications: ¿none¿; condition of patient: ¿stable.".
|
|
Manufacturer Narrative
|
Device code(s): appropriate term/code not available (3191) was selected for the vena cava perforation.Investigation ¿ investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Vena cava wall penetration/perforation has been reported and may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Unknown if the reported mental and physical pain and permanent impairment are directly related to the filter and unable to identify a corresponding failure mode at this time.The following allegations have been investigated: vena cava perforation, mental and physical pain, permanent impairment.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.Blank fields on this form indicate the information is unknown or unavailable, or unchanged.
|
|
Event Description
|
The patient reportly expired.Specific details surrounding the patient's expiration are currently unknown, and there is currently no reported allegation of wrongful death.
|
|
Manufacturer Narrative
|
Investigation is reopened due to additional information provided.Per quality engineering review, the additional information provided for this complaint does not change the previous investigation conclusion.Therefore, no new investigation activities will be conducted at this time.Cook will reopen the investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.20 devices in lot.No relevant notes on wo for neither device (igtcfs-65-1-fem-tulip) lot: e2792091, nor filter (igtf-30) lot: e2785815 and e2784937.No other complaints on lots.Product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|