• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP CONSOLE; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP CONSOLE; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE Back to Search Results
Model Number 60-03-75
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Patient Involvement (2645)
Event Date 10/22/2018
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump console.The incident occurred in (b)(6).The livanova field representative who performed the maintenance checked all the modules, reconnected all the cables, but the issue persisted.Livanova (b)(4) requested the entire unit back for a detailed investigation but it was not made available.No investigation could not be carried out and no root cause could be determined.A review of the dhr could not identify any deviations or nonconformities relevant to the issue.
 
Event Description
Livanova received a report that, during a battery test, the message "no com" was displayed.The issue occurred during maintenance.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP CONSOLE
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 89309
GM  89309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 89309
GM   89309
Manufacturer Contact
barbara galasso
14401 w. 65th way
arvada, CO 80004
MDR Report Key8748004
MDR Text Key149810040
Report Number9611109-2019-00503
Device Sequence Number1
Product Code KFM
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K020571
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 07/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/01/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-03-75
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/22/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/23/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-