Model Number 439888
Device Problems Failure to Capture (1081); Device Dislodged or Dislocated (2923)
Patient Problem Ventricular Tachycardia (2132)
Event Date 01/22/2019
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that there was failure to capture and dislodgement on the left ventricular (lv) lead.It was noted the lv lead appeared to be in the right ventricular (rv) and is likely causing mechanical irritation of the myocardium, which led to 2 episodes of ventricular tachycardia (vt).The lead was explanted and replaced.The patient is a participant in the post approval clinical surveillance product surveillance registry.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN PERFORMA STRAIGHT MRI SURESCAN
Type of Device
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key8757552
MDR Text Key149924361
Report Number2649622-2019-11470
Device Sequence Number1
Product Code OJX
UDI-Device Identifier00643169601857
UDI-Public00643169601857
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P080006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 07/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/03/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/12/2020
Device Model Number439888
Device Catalogue Number439888
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/12/2019
Date Device Manufactured12/07/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
5076-52 LEAD, W4TR01 IPG, 6725 ADAPTOR
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age75 YR
Patient Weight101
-
-