Model Number L311
Device Problems Failure to Capture (1081); High Capture Threshold (3266)
Patient Problem Injury (2348)
Event Date 05/27/2019
Event Type  Injury  
Manufacturer Narrative
At this time, the product has not been returned.If the product is returned, analysis will be performed and this report would be updated at that time.
 
Event Description
It was reported that the patient implanted with this pacemaker and right ventricular (rv) lead was hospitalized due to confirmed asystole due to intermittent loss of capture (loc) because of high thresholds.The patient is pacemaker dependent.Surgical intervention was performed.The device was explanted and the rv lead was capped.
 
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.A visual inspection of the device header and case noted no anomalies.Pin gauge testing, designed to verify proper port dimensions, was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed, and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.
 
Event Description
The product has been received for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE MRI DR
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key8758895
MDR Text Key149971705
Report Number2124215-2019-12421
Device Sequence Number1
Product Code LWP
UDI-Device Identifier00802526559228
UDI-Public00802526559228
Combination Product (y/n)N
PMA/PMN Number
P150012/S000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/03/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/06/2017
Device Model NumberL311
Device Catalogue NumberL311
Device Lot Number103010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/19/2019
Date Manufacturer Received10/18/2019
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age70 YR
-
-