STRYKER ORTHOPAEDICS-MAHWAH TRI TS BASEPLATE SIZE 3; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, C
|
Back to Search Results |
|
Catalog Number 5521-B-300 |
Device Problems
Manufacturing, Packaging or Shipping Problem (2975); Packaging Problem (3007)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/12/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
A review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
|
|
Event Description
|
Primary procedure, right tka.It was reported that when the outer blister of the tibial baseplate was opened, the tyvek of the inner blister was observed to be already opened (not sealed all the way around).Due to sterility concerns, the device was not used.Another device was immediately available in the operating room.Rep confirmed that surgery was completed successfully with no delay, and that no further information is available.
|
|
Event Description
|
Primary procedure, right tka.It was reported that when the outer blister of the tibial baseplate was opened, the tyvek of the inner blister was observed to be already opened (not sealed all the way around).Due to sterility concerns, the device was not used.Another device was immediately available in the o.R.Rep confirmed that surgery was completed successfully with no delay and that no further information is available.
|
|
Manufacturer Narrative
|
Reported as serious injury in error.An event regarding a packaging issue involving a triathlon baseplate was reported.The event was confirmed upon visual inspection of the device returned.Method & results: device evaluation and results: device arrived without carton and outer tyvek lid.Inner tyvek lid has a discontinuity in one corner of the inner blister seal.Inner tyvek lid also has a crease line, indicating that it has been folded back at this corner.Clinician review: no medical records were received for review with a clinical consultant.Device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.No further investigation for this event is possible at this time as insufficient information was received by stryker orthopaedics.If the secondary packaging and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
|